Cloud-clone
SKU:LMB785Mu
Magnetic Luminex Assay Kit for Neprilysin (CD10) ,etc.
Magnetic Luminex Assay Kit for Neprilysin (CD10) ,etc.
Couldn't load pickup availability
Product No.
LMB785Mu
Organism Species
Mus musculus (Mouse).
Sample Type
Serum, plasma and other biological fluids.
Test Method
Double-antibody Sandwich
Assay Length
3.5h
Detection Range
0.98-1000pg/mL
Sensitivity
The minimum detectable dose of this kit is typically less than 0.327 pg/mL.
UOM
1Plex 2Plex 3Plex 4Plex 5Plex 6Plex 7Plex 8Plex
Specificity
This assay has high sensitivity and excellent specificity for detection of Neprilysin (CD10) ,etc..
No significant cross-reactivity or interference between Neprilysin (CD10) ,etc. and analogues was observed.
Recovery
Matrices listed below were spiked with certain level of recombinant Neprilysin (CD10) ,etc. and the recovery rates were calculated by comparing the measured value to the expected amount of Neprilysin (CD10) ,etc. in samples.
| Matrix | Recovery range (%) | Average(%) |
| serum(n=5) | 87-105 | 96 |
| EDTA plasma(n=5) | 88-96 | 91 |
| heparin plasma(n=5) | 78-93 | 85 |
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, middle and high level Neprilysin (CD10) ,etc. were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, middle and high level Neprilysin (CD10) ,etc. were tested on 3 different plates, 8 replicates in each plate.
CV(%) = SD/meanX100
Intra-Assay: CV<10%
Inter-Assay: CV<12%
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Neprilysin (CD10) ,etc. and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
| Sample | 1:2 | 1:4 | 1:8 | 1:16 |
| serum(n=5) | 81-96% | 83-94% | 80-95% | 78-101% |
| EDTA plasma(n=5) | 88-96% | 87-98% | 97-104% | 95-103% |
| heparin plasma(n=5) | 78-90% | 80-90% | 96-103% | 88-102% |
Stability
The stability of kit is determined by the loss rate of activity. The loss rate of this kit is less than 5% within the expiration date under appropriate storage condition.
To minimize extra influence on the performance, operation procedures and lab conditions, especially room temperature, air humidity, incubator temperature should be strictly controlled. It is also strongly suggested that the whole assay is performed by the same operator from the beginning to the end.
Reagents and materials provided
| Reagents | Quantity | Reagents | Quantity |
| 96-well plate | 1 | Plate sealer for 96 wells | 4 |
| Pre-Mixed Standard | 2 | Standard Diluent | 1×20mL |
| Pre-Mixed Magnetic beads (22#:CD10) | 1 | Analysis buffer | 1×20mL |
| Pre-Mixed Detection Reagent A | 1×120μL | Assay Diluent A | 1×12mL |
| Detection Reagent B (PE-SA) | 1×120μL | Assay Diluent B | 1×12mL |
| Sheath Fluid | 1×10mL | Wash Buffer (30 × concentrate) | 1×20mL |
| Instruction manual | 1 |
Assay procedure summary
1. Preparation of standards, reagents and samples before the experiment;
2. Add 100μL standard or sample to each well,
add 10μL magnetic beads, and incubate 90min at 37°C on shaker;
3. Remove liquid on magnetic frame, add 100μL prepared Detection Reagent A. Incubate 60min at 37°C on shaker;
4. Wash plate on magnetic frame for three times;
5. Add 100μL prepared Detection Reagent B, and incubate 30 min at 37°C on shaker;
6. Wash plate on magnetic frame for three times;
7. Add 100μL sheath solution, swirl for 2 minutes, read on the machine.
| Magazine | Citations |
| 13 | Elevated Plasma levels of Neutral Endpeptidase in Hypertension Patients Treated with Angiotensin Converting Enzyme Inhibitors Ahajournals:Source |
| Biol Pharm Bull | Effect of High Glucose Levels on Amyloid β Production in Retinas of Spontaneous Diabetes Mellitus Otsuka Long-Evans Tokushima Fatty Rats PubMed: 25832640 |
| Eur J Heart Fail. | Soluble neprilysin does not correlate with outcome in heart failure with preserved ejection fraction Pubmed:26725876 |
| 8 | A Test in Context: Neprilysin: Function, Inhibition, and Biomarker. pubmed:27491909 |
| European Journal of Heart Failure | Effects of sacubitril/valsartan on neprilysin targets and the metabolism of natriuretic peptides in chronic heart failure: a mechanistic clinical study Pubmed: 30520545 |
| European Journal of Heart Failure | Increased granulocyte membrane neprilysin (CD10) expression is associated with better prognosis in heart failure Pubmed: 30828920 |
| Clinical Research in Cardiology | Soluble neprilysin, NT-proBNP, and growth differentiation factor-15 as biomarkers for heart failure in dialysis patients (SONGBIRD) Pubmed: 32002632 |
| Neurotoxicology | Involvement of hippocampal agmatine in β1-42 amyloid induced memory impairment, neuroinflammation and BDNF signaling disruption in mice Pubmed: 32522471 |
| European Heart Journal | P1620 Association of granulocyte neprilysin (CD10) expression with prognosis in heart failure with reduced ejection fraction |
| Clin Biochem | Finding a reliable assay for soluble neprilysin Pubmed:35331754 |
| Scientific Reports | The potential role of serum tau protein (MAPT), neuronal cell adhesion molecule (NrCAM) and neprilysin (NEP) in neurodegenerative disorders development in … |
| Catalog No. | Related products for research use of Mus musculus (Mouse) Organism species | Applications (RESEARCH USE ONLY!) |
| RPB785Mu01 | Recombinant Neprilysin (CD10) | Positive Control; Immunogen; SDS-PAGE; WB. |
| RPB785Mu02 | Recombinant Neprilysin (CD10) | Positive Control; Immunogen; SDS-PAGE; WB. |
| PAB785Mu01 | Polyclonal Antibody to Neprilysin (CD10) | WB; IHC; ICC; IP. |
| PAB785Mu02 | Polyclonal Antibody to Neprilysin (CD10) | WB; IHC; ICC; IP. |
| SEB785Mu | ELISA Kit for Neprilysin (CD10) | Enzyme-linked immunosorbent assay for Antigen Detection. |
| LMB785Mu | Magnetic Luminex Assay Kit for Neprilysin (CD10) ,etc. | Magnetic Luminex Assay for Antigen Detection. |