FineTest
SKU:ECH0035
Chicken IL-13(Interleukin 13) ELISA Kit
Chicken IL-13(Interleukin 13) ELISA Kit
Couldn't load pickup availability
The Chicken IL-13 (Interleukin 13) ELISA Kit is a specialized research tool designed for the quantitative measurement of Interleukin 13 levels in chicken samples. IL-13 is a critical cytokine involved in immune responses and inflammation regulation. This kit empowers researchers and clinicians to accurately assess IL-13 levels in chicken biological samples. Key Features: High Specificity: This kit is specifically designed to detect chicken IL-13, ensuring minimal cross-reactivity and accurate results. Quantitative Measurements: Provides precise quantitative data for comparisons and data analysis. Versatile Application: Suitable for use in various chicken sample types, accommodating a wide range of research and clinical applications. User-Friendly: The kit includes all necessary components and detailed instructions for straightforward and efficient experimentation. Reliability: Stringent quality control processes and calibration standards ensure the reliability and consistency of results. Advance your research in avian immunology, inflammation regulation, and cytokine biology with the Chicken IL-13 (Interleukin 13) ELISA Kit.
Product Name
Chicken IL-13(Interleukin 13) ELISA Kit
Alias
IL-13 ELISA Kit, Interleukin 13 ELISA Kit, BHR1interleukin-13 ELISA Kit
Detection Method
Sandwich ELISA, Double Antibody
Application
IL-13 ELISA Kit allows for the in vitro quantitative determination of IL-13 concentrations in serum, plasma, tissue homogenates and other biological fluids.
Size
96T
Range
15.625-1000pg/ml
Sensitivity
< 9.375pg/ml
Species
Chicken
UniProt ID
Standard Curve

Storage
2-8 ℃ for 6 months
Recovery
Matrices listed below were spiked with certain level of IL-13 and the recovery rates were calculated by comparing the measured value to the expected amount of IL-13 in samples.
|
Matrix |
Recovery range(%) |
Average(%) |
|
serum(n=5) |
87-102 |
94 |
|
EDTA plasma(n=5) |
88-105 |
94 |
|
heparin plasma(n=5) |
92-102 |
96 |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of IL-13 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
|
Sample |
1:2 |
1:4 |
1:8 |
|
serum(n=5) |
86-103% |
85-96% |
91-105% |
|
EDTA plasma(n=5) |
84-100% |
83-100% |
86-89% |
|
heparin plasma(n=5) |
85-99% |
80-95% |
82-91% |
CV(%)
Intra-Assay: CV<8%
Inter-Assay: CV<10%
Note
For Research Use Only
Share