Overview
| Product name: | Human TG-ab(anti-Thyroglobulin Ab) ELISA Kit |
| Reactivity: | Human |
| Alternative Names: | TG AB; TG-AB |
| Assay Type: | Competitive Inhibition |
| Sensitivity: | 0.51 ng/mL |
| Standard: | 100 ng/mL |
| Detection Range: | 1.57-100 ng/mL |
| Sample Type: | serum, plasma, cell lysates, cell culture supernates and other biological fluids |
| Assay Length: | 2h |
| Research Area: | Signal transduction;Metabolic pathway;Infection immunity;Hormone metabolism;Dermatology; |
| Test principle: | This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with Human TG-Ab protein. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human TG-Ab. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human TG-Ab in the samples is then determined by comparing the OD of the samples to the standard curve. |
Standard curve
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 100.00 | 0.227 | — |
| 50.00 | 0.386 | — |
| 25.00 | 0.621 | — |
| 12.50 | 0.861 | — |
| 6.25 | 1.168 | — |
| 3.13 | 1.557 | — |
| 1.57 | 1.796 | — |
| 0.00 | 2.115 | — |

Precision
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Recovery
Matrices listed below were spiked with certain level of recombinant TG-Ab and the recovery rates were calculated by comparing the measured value to the expected amount of TG-Ab in samples.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 78-90% | 84% |
| EDTA plasma(n=5) | 85-99% | 92% |
| Heparin plasma(n=5) | 90-105% | 97% |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of TG-Ab and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-101% | 89-102% | 88-103% | 83-94% |
| EDTA plasma(n=5) | 92-105% | 89-97% | 87-95% | 83-96% |
| Heparin plasma(n=5) | 92-102% | 82-93% | 83-97% | 92-103% |
Share