1
/
of
1
ELK Biotechnology
SKU:ELK9197
Human Pro-ADM(Proadrenomedullin) ELISA Kit
Human Pro-ADM(Proadrenomedullin) ELISA Kit
Regular price
$458.00 USD
Regular price
Sale price
$458.00 USD
Unit price
/
per
Shipping calculated at checkout.
Couldn't load pickup availability
Human Pro-ADM(Proadrenomedullin) ELISA Kit
Overview
| Product name: | Human Pro-ADM(Proadrenomedullin) ELISA Kit |
| Reactivity: | Human |
| Alternative Names: | Proadrenomedullin |
| Assay Type: | Sandwich |
| Sensitivity: | 0.114 ng/mL |
| Standard: | 20 ng/mL |
| Detection Range: | 0.32-20 ng/mL |
| Sample Type: | tissue homogenates, cell lysates, cell culture supernates and other biological fluids |
| Assay Length: | 3.5h |
| Research Area: | Signal transduction; |
| Test principle: | The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human Pro-ADM. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human Pro-ADM. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human Pro-ADM, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human Pro-ADM in the samples is then determined by comparing the OD of the samples to the standard curve. |
Standard curve
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 20.00 | 2.204 | 2.117 |
| 10.00 | 1.628 | 1.541 |
| 5.00 | 1.182 | 1.095 |
| 2.50 | 0.835 | 0.748 |
| 1.25 | 0.473 | 0.386 |
| 0.63 | 0.305 | 0.218 |
| 0.32 | 0.215 | 0.128 |
| 0.00 | 0.087 | 0.000 |

Precision
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Recovery
Matrices listed below were spiked with certain level of recombinant Pro-ADM and the recovery rates were calculated by comparing the measured value to the expected amount of Pro-ADM in samples.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 86-99% | 92% |
| EDTA plasma(n=5) | 87-99% | 93% |
| Heparin plasma(n=5) | 81-97% | 89% |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Pro-ADM and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 81-96% | 82-95% | 86-99% | 89-96% |
| EDTA plasma(n=5) | 93-107% | 93-102% | 79-94% | 87-98% |
| Heparin plasma(n=5) | 86-97% | 90-101% | 92-102% | 81-93% |