Overview
| Product name: | Human sTREM-1(soluble Triggering Receptor Expressed on Myeloid Cells-1) ELISA Kit |
| Reactivity: | Human |
| Alternative Names: | CD354; Triggering Receptor Expressed On Monocytes-1; sTREM1 |
| Assay Type: | Sandwich |
| Sensitivity: | 13.1 pg/mL |
| Standard: | 2000 pg/mL |
| Detection Range: | 31.3-2000 pg/mL |
| Sample Type: | serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids |
| Assay Length: | 3.5h |
| Research Area: | Immunology |
| Test principle: | The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human sTREM-1. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human sTREM-1. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human sTREM-1, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human sTREM-1 in the samples is then determined by comparing the OD of the samples to the standard curve. |
Standard curve
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 2.135 | 2.031 |
| 1000.00 | 1.597 | 1.493 |
| 500.00 | 1.185 | 1.081 |
| 250.00 | 0.963 | 0.859 |
| 125.00 | 0.492 | 0.388 |
| 62.50 | 0.355 | 0.251 |
| 31.25 | 0.208 | 0.104 |
| 0.00 | 0.104 | 0.000 |

Precision
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Recovery
Matrices listed below were spiked with certain level of recombinant sTREM-1 and the recovery rates were calculated by comparing the measured value to the expected amount of sTREM-1 in samples.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 79-97% | 88% |
| EDTA plasma(n=5) | 90-105% | 97% |
| Heparin plasma(n=5) | 82-97% | 89% |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of sTREM-1 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 79-92% | 89-102% | 87-98% | 80-92% |
| EDTA plasma(n=5) | 98-105% | 92-101% | 85-105% | 92-104% |
| Heparin plasma(n=5) | 82-96% | 95-102% | 89-97% | 95-106% |