1
/
of
1
ELK Biotechnology
SKU:ELK9087
Cattle ADMA(Asymmetrical Dimethylarginine) ELISA Kit
Cattle ADMA(Asymmetrical Dimethylarginine) ELISA Kit
Regular price
$376.00 USD
Regular price
Sale price
$376.00 USD
Unit price
/
per
Shipping calculated at checkout.
Couldn't load pickup availability
Cattle ADMA(Asymmetrical Dimethylarginine) ELISA Kit
Overview
| Product name: | Cattle ADMA(Asymmetrical Dimethylarginine) ELISA Kit |
| Reactivity: | Cattle |
| Alternative Names: | N,N-Dimethylarginine; NG,NG-Dimethylarginine dihydrochloride; Asymmetric Dimethylarginine |
| Assay Type: | Competitive Inhibition |
| Sensitivity: | 4.51 ng/mL |
| Standard: | 1000 ng/mL |
| Detection Range: | 15.63-1000 ng/mL |
| Sample Type: | serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids |
| Assay Length: | 2h |
| Research Area: | Metabolic pathway;Cardiovascular biology; |
| Test principle: | This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with Cattle ADMA protein. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Cattle ADMA. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Cattle ADMA in the samples is then determined by comparing the OD of the samples to the standard curve. |
Standard curve
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 0.197 | — |
| 500.00 | 0.413 | — |
| 250.00 | 0.611 | — |
| 125.00 | 0.842 | — |
| 62.50 | 1.253 | — |
| 31.25 | 1.537 | — |
| 15.63 | 1.762 | — |
| 0.00 | 2.256 | — |

Precision
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Recovery
Matrices listed below were spiked with certain level of recombinant ADMA and the recovery rates were calculated by comparing the measured value to the expected amount of ADMA in samples.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 84-92% | 87% |
| EDTA plasma(n=5) | 83-95% | 89% |
| Heparin plasma(n=5) | 88-102% | 85% |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of ADMA and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 93-101% | 87-98% | 83-95% | 88-95% |
| EDTA plasma(n=5) | 85-102% | 82-96% | 87-98% | 78-92% |
| Heparin plasma(n=5) | 87-101% | 93-105% | 92-104% | 82-96% |
Share