Product name: Hamster D2D(D-Dimer) ELISA Kit
Reactivity: Hamster
Alternative Names: D 2 Dimer
Assay Type: Competitive Inhibition
Sensitivity: 5.52 ng/mL
Standard: 1000 ng/mL
Detection Range: 15.63-1000 ng/mL
Sample Type: plasma
Assay Length: 2h
Research Area: Hematology;Cardiovascular biology;
Test principle: This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with Hamster Hamster D2D protein. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Hamster Hamster D2D. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Hamster Hamster D2D in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (ng/mL) OD Corrected OD
1000.00 0.191
500.00 0.357
250.00 0.655
125.00 0.842
62.50 1.127
31.25 1.511
15.63 1.871
0.00 2.115

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant Hamster D2D and the recovery rates were calculated by comparing the measured value to the expected amount of Hamster D2D in samples.
Matrix Recovery range Average
serum(n=5) 92-107% 99%
EDTA plasma(n=5) 78-95% 86%
Heparin plasma(n=5) 87-95% 98%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Hamster D2D and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 92-96% 86-98% 88-104% 87-93%
EDTA plasma(n=5) 86-95% 82-93% 97-106% 89-102%
Heparin plasma(n=5) 81-94% 81-96% 93-106% 82-95%