Overview
| Product name: | Human sTNF-R1(soluble tumor necrosis factor receptor 1) ELISA Kit |
| Reactivity: | Human |
| Alternative Names: | sTNF-R1; soluble tumor necrosis factor receptor 1 |
| Assay Type: | Sandwich |
| Sensitivity: | 3.9 pg/mL |
| Standard: | 1000 pg/mL |
| Range: | 15.63-1000 pg/mL |
| Sample Type: | serum, plasma and other biological fluids |
| Assay Length: | 3.5h |
| Research Area: | Cancer |
| Test principle: | The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human sTNF-R1. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human sTNF-R1. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human sTNF-R1, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human sTNF-R1 in the samples is then determined by comparing the OD of the samples to the standard curve. |
Standard curve
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 1.916 | 1.837 |
| 500.00 | 1.643 | 1.564 |
| 250.00 | 1.107 | 1.028 |
| 125.00 | 0.852 | 0.773 |
| 62.50 | 0.516 | 0.437 |
| 31.25 | 0.381 | 0.302 |
| 15.63 | 0.237 | 0.158 |
| 0.00 | 0.079 | 0.000 |

Precision
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Recovery
Matrices listed below were spiked with certain level of recombinant sTNF-R1 and the recovery rates were calculated by comparing the measured value to the expected amount of sTNF-R1 in samples.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-105% | 97% |
| EDTA plasma(n=5) | 97-105% | 101% |
| Heparin plasma(n=5) | 95-107% | 101% |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of sTNF-R1 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 79-91% | 87-96% | 83-101% | 82-95% |
| EDTA plasma(n=5) | 85-93% | 87-96% | 97-105% | 95-102% |
| Heparin plasma(n=5) | 83-92% | 96-102% | 81-93% | 79-91% |
Share