Overview
| Product name: | Mouse 4-Hydroxynonenal(HNE) ELISA Kit |
| Reactivity: | Mouse |
| Alternative Names: | 4-HNE, HNE, 4-hydroxy-2-nonenal, 6-Hydroxynonenal |
| Assay Type: | Sandwich |
| Sensitivity: | 1.12 ng/mL |
| Standard: | 150 ng/mL |
| Range: | 2.35-150 ng/mL |
| Sample Type: | Serum, plasma and other biological fluids. |
| Assay Length: | 3.5h |
| Research Area: | Signal transduction;Metabolic pathway;Infection immunity;Hormone metabolism;Dermatology; |
| Test principle: | The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Mouse 4-HNE. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Mouse 4-HNE. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Mouse 4-HNE, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Mouse 4-HNE in the samples is then determined by comparing the OD of the samples to the standard curve. |
Standard curve
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 150.00 | 2.035 | 1.946 |
| 75.00 | 1.782 | 1.693 |
| 37.50 | 1.297 | 1.208 |
| 18.75 | 0.883 | 0.794 |
| 9.38 | 0.589 | 0.500 |
| 4.69 | 0.327 | 0.238 |
| 2.35 | 0.197 | 0.108 |
| 0.00 | 0.089 | 0.000 |

Precision
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Recovery
Matrices listed below were spiked with certain level of recombinant 4-HNE and the recovery rates were calculated by comparing the measured value to the expected amount of 4-HNE in samples.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 82-94% | 88% |
| EDTA plasma(n=5) | 84-98% | 91% |
| Heparin plasma(n=5) | 80-98% | 89% |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of 4-HNE and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 78-92% | 96-105% | 86-97% | 85-101% |
| EDTA plasma(n=5) | 78-91% | 86-92% | 89-96% | 91-97% |
| Heparin plasma(n=5) | 95-104% | 97-105% | 82-95% | 89-97% |