1
/
of
1
ELK Biotechnology
SKU:ELK8127
Dog TNNT2(Troponin T Type 2, Cardiac) ELISA Kit
Dog TNNT2(Troponin T Type 2, Cardiac) ELISA Kit
Regular price
$588.00 USD
Regular price
Sale price
$588.00 USD
Unit price
/
per
Shipping calculated at checkout.
Couldn't load pickup availability
Dog TNNT2(Troponin T Type 2, Cardiac) ELISA Kit
Overview
| Product name: | Dog cTnT/TNNT2(Troponin T Type 2, Cardiac) ELISA Kit |
| Reactivity: | Dog |
| Alternative Names: | CMH2; cTnT; TnTc; Cardiomyopathy,Hypertrophic 2; Cardiac muscle troponin T; TNNT2; Troponin T Type 2, Cardiac |
| Assay Type: | Sandwich |
| Sensitivity: | 23.9 pg/mL |
| Standard: | 4000 pg/mL |
| Range: | 62.5-4000 pg/mL |
| Sample Type: | Serum, plasma, tissue homogenates and other biological fluids |
| Assay Length: | 3.5h |
| Research Area: | Cardiovascular biology; |
| Test principle: | The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Dog cTnT/TNNT2. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Dog cTnT/TNNT2. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Dog cTnT/TNNT2, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Dog cTnT/TNNT2 in the samples is then determined by comparing the OD of the samples to the standard curve. |
Standard curve
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 4000.00 | 2.268 | 2.182 |
| 2000.00 | 1.562 | 1.476 |
| 1000.00 | 1.067 | 0.981 |
| 500.00 | 0.872 | 0.786 |
| 250.00 | 0.529 | 0.443 |
| 125.00 | 0.385 | 0.299 |
| 62.50 | 0.206 | 0.120 |
| 0.00 | 0.086 | 0.000 |

Precision
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Recovery
Matrices listed below were spiked with certain level of recombinant cTnT/TNNT2 and the recovery rates were calculated by comparing the measured value to the expected amount of cTnT/TNNT2 in samples.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 88-101% | 94% |
| EDTA plasma(n=5) | 82-95% | 88% |
| Heparin plasma(n=5) | 78-91% | 84% |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of cTnT/TNNT2 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-102% | 85-94% | 87-101% | 89-99% |
| EDTA plasma(n=5) | 86-101% | 86-101% | 87-96% | 92-105% |
| Heparin plasma(n=5) | 89-102% | 92-101% | 88-102% | 87-102% |
Share