1
/
of
1
ELK Biotechnology
SKU:ELK8059
General 5-Me-THF(5-Methyltetrahydrofolate) ELISA Kit
General 5-Me-THF(5-Methyltetrahydrofolate) ELISA Kit
Regular price
$376.00 USD
Regular price
Sale price
$376.00 USD
Unit price
/
per
Shipping calculated at checkout.
Couldn't load pickup availability
General 5-Me-THF(5-Methyltetrahydrofolate) ELISA Kit
Overview
| Product name: | 5-Me-THF(5-Methyltetrahydrofolate) ELISA Kit |
| Reactivity: | General |
| Alternative Names: | 5-Me-THFA; 5-Me-H4FA; 5-Me-THF; 5-Me-H4F; Levomefolic Acid; Metafolin; 5-Methyltetrahydrofolic Acid; L-Methylfolate |
| Assay Type: | Competitive Inhibition |
| Sensitivity: | 0.43 ng/mL |
| Standard: | 90 ng/mL |
| Detection Range: | 1.41-90 ng/mL |
| Sample Type: | serum, plasma and other biological fluids |
| Assay Length: | 2h |
| Research Area: | Hematology;Reproductive science;Nutrition metabolism; |
| Test principle: | This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with 5-Me-THF protein. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to 5-Me-THF. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of 5-Me-THF in the samples is then determined by comparing the OD of the samples to the standard curve. |
Standard curve
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 90.00 | 0.197 | — |
| 45.00 | 0.357 | — |
| 22.50 | 0.611 | — |
| 11.25 | 0.812 | — |
| 5.63 | 1.197 | — |
| 2.82 | 1.511 | — |
| 1.41 | 1.796 | — |
| 0.00 | 2.213 | — |

Precision
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Recovery
Matrices listed below were spiked with certain level of recombinant 5-Me-THF and the recovery rates were calculated by comparing the measured value to the expected amount of 5-Me-THF in samples.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 92-106% | 99% |
| EDTA plasma(n=5) | 77-95% | 86% |
| Heparin plasma(n=5) | 88-103% | 91% |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of 5-Me-THF and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-104% | 84-98% | 92-103% | 82-96% |
| EDTA plasma(n=5) | 86-97% | 83-96% | 91-103% | 96-105% |
| Heparin plasma(n=5) | 88-103% | 90-98% | 92-106% | 85-97% |
Share