Overview
| Product name: | Human Anti-IgA(Anti-Immunoglobulin A Antibody) ELISA Kit |
| Reactivity: | Human |
| Alternative Names: | anti-IgA |
| Assay Type: | Sandwich |
| Sensitivity: | 2.36 ng/mL |
| Standard: | 400 ng/mL |
| Detection Range: | 6.25-400 ng/mL |
| Sample Type: | serum, plasma and other biological fluids |
| Assay Length: | 3.5h |
| Research Area: | Infection immunity;Immune molecule;Hematology;Pulmonology; |
| Test principle: | The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human Anti-IgA. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human Anti-IgA. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human Anti-IgA, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human Anti-IgA in the samples is then determined by comparing the OD of the samples to the standard curve. |
Standard curve
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 400.00 | 2.045 | 1.964 |
| 200.00 | 1.527 | 1.446 |
| 100.00 | 1.088 | 1.007 |
| 50.00 | 0.892 | 0.811 |
| 25.00 | 0.547 | 0.466 |
| 12.50 | 0.403 | 0.322 |
| 6.25 | 0.179 | 0.098 |
| 0.00 | 0.081 | 0.000 |

Precision
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Recovery
Matrices listed below were spiked with certain level of recombinant Anti-IgA and the recovery rates were calculated by comparing the measured value to the expected amount of Anti-IgA in samples.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-97% | 91% |
| EDTA plasma(n=5) | 82-94% | 88% |
| Heparin plasma(n=5) | 94-105% | 98% |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Anti-IgA and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-94% | 87-101% | 86-97% | 88-95% |
| EDTA plasma(n=5) | 96-103% | 83-96% | 87-98% | 85-97% |
| Heparin plasma(n=5) | 78-93% | 96-102% | 95-103% | 89-97% |
Share