Skip to product information
1 of 1

ELK Biotechnology

SKU:ELK7450

Rat MSTN(Myostatin) ELISA Kit

Rat MSTN(Myostatin) ELISA Kit

Regular price $458.00 USD
Regular price Sale price $458.00 USD
Sale Sold out
Shipping calculated at checkout.
Size
Sensitivity
Detection range

Rat MSTN(Myostatin) ELISA Kit

Overview

Product name: Rat MSTN(Myostatin) ELISA Kit
Reactivity: Rat
Alternative Names: GDF8; Growth Differentiation Factor 8
Assay Type: Competitive Inhibition
Sensitivity: 0.44 ng/mL
Standard: 90 ng/mL
Range: 1.41-90 ng/mL
Sample Type: serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Length: 2h
Research Area: Cytokine;Metabolic pathway;Bone metabolism;
Test principle: This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with Rat MSTN protein. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Rat MSTN. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Rat MSTN in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (ng/mL) OD Corrected OD
90.00 0.195
45.00 0.479
22.50 0.786
11.25 0.967
5.63 1.236
2.82 1.596
1.41 1.877
0.00 2.232

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant MSTN and the recovery rates were calculated by comparing the measured value to the expected amount of MSTN in samples.
Matrix Recovery range Average
serum(n=5) 95-106% 101%
EDTA plasma(n=5) 84-99% 91%
Heparin plasma(n=5) 87-102% 93%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of MSTN and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 88-97% 96-103% 89-97% 91-103%
EDTA plasma(n=5) 85-94% 87-101% 86-97% 88-95%
Heparin plasma(n=5) 95-103% 86-92% 81-93% 79-87%