1
/
of
1
ELK Biotechnology
SKU:ELK7048
Rat CDH5(Cadherin 5) ELISA Kit
Rat CDH5(Cadherin 5) ELISA Kit
Regular price
$458.00 USD
Regular price
Sale price
$458.00 USD
Unit price
/
per
Shipping calculated at checkout.
Couldn't load pickup availability
Rat CDH5(Cadherin 5) ELISA Kit
Overview
| Product name: | Rat VE-Cadherin(Vascular Endothelial Cadherin) ELISA Kit |
| Reactivity: | Rat |
| Alternative Names: | CD144; 7B4; Cadherin 5 Type 2; VE-Cadherin; Cadherin,Vascular Endothelial; CDH5; Cadherin 5 |
| Assay Type: | Sandwich |
| Sensitivity: | 27 pg/mL |
| Standard: | 5000 pg/mL |
| Range: | 78.13-5000 pg/mL |
| Sample Type: | serum, plasma, tissue homogenates, cell lysates and other biological fluids |
| Assay Length: | 3.5h |
| Research Area: | CD & Adhesion molecule;Tumor immunity;Infection immunity; |
| Test principle: | The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Rat VE-Cadherin. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Rat VE-Cadherin. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Rat VE-Cadherin, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Rat VE-Cadherin in the samples is then determined by comparing the OD of the samples to the standard curve. |
Standard curve
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 5000.00 | 1.946 | 1.850 |
| 2500.00 | 1.527 | 1.431 |
| 1250.00 | 1.088 | 0.992 |
| 625.00 | 0.682 | 0.586 |
| 312.50 | 0.539 | 0.443 |
| 156.25 | 0.305 | 0.209 |
| 78.13 | 0.249 | 0.153 |
| 0.00 | 0.096 | 0.000 |

Precision
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Recovery
Matrices listed below were spiked with certain level of recombinant VE-Cadherin and the recovery rates were calculated by comparing the measured value to the expected amount of VE-Cadherin in samples.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 82-95% | 88% |
| EDTA plasma(n=5) | 95-107% | 101% |
| Heparin plasma(n=5) | 94-107% | 100% |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of VE-Cadherin and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-95% | 85-98% | 82-97% | 92-101% |
| EDTA plasma(n=5) | 84-94% | 85-103% | 94-105% | 92-101% |
| Heparin plasma(n=5) | 81-94% | 96-105% | 92-103% | 82-96% |
Share