1
/
of
1
ELK Biotechnology
SKU:ELK5403
Mouse PINP(Procollagen I N-Terminal Propeptide) ELISA Kit
Mouse PINP(Procollagen I N-Terminal Propeptide) ELISA Kit
Regular price
$458.00 USD
Regular price
Sale price
$458.00 USD
Unit price
/
per
Shipping calculated at checkout.
Couldn't load pickup availability
Mouse PINP(Procollagen I N-Terminal Propeptide) ELISA Kit
Overview
| Product name: | Mouse PⅠNP(Procollagen Ⅰ N-Terminal Propeptide) ELISA Kit |
| Reactivity: | Mouse |
| Alternative Names: | P1NP; N-Propeptide Of Type I Procollagen; Procollagen I Amino Terminal Propeptide; PINP; Procollagen I N-Terminal Propeptide |
| Assay Type: | Competitive Inhibition |
| Sensitivity: | 152.5 pg/mL |
| Standard: | 30000 pg/mL |
| Range: | 468.75-30000 pg/mL |
| Sample Type: | serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids |
| Assay Length: | 2h |
| Research Area: | Metabolic pathway;Hepatology;Bone metabolism; |
| Test principle: | This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with Mouse PⅠNP protein. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Mouse PⅠNP. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Mouse PⅠNP in the samples is then determined by comparing the OD of the samples to the standard curve. |
Standard curve
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 30000.00 | 0.179 | — |
| 15000.00 | 0.397 | — |
| 7500.00 | 0.527 | — |
| 3750.00 | 0.897 | — |
| 1875.00 | 1.297 | — |
| 937.50 | 1.457 | — |
| 468.75 | 1.863 | — |
| 0.00 | 2.232 | — |

Precision
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Recovery
Matrices listed below were spiked with certain level of recombinant PⅠNP and the recovery rates were calculated by comparing the measured value to the expected amount of PⅠNP in samples.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 85-97% | 91% |
| EDTA plasma(n=5) | 96-108% | 102% |
| Heparin plasma(n=5) | 80-95% | 89% |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of PⅠNP and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 90-99% | 83-97% | 95-102% | 87-98% |
| EDTA plasma(n=5) | 78-92% | 79-93% | 94-106% | 83-97% |
| Heparin plasma(n=5) | 89-97% | 95-104% | 82-96% | 85-98% |
Share