Skip to product information
1 of 1

ELK Biotechnology

SKU:ELK5108

Mouse HbA1c(Glycated Hemoglobin A1c) ELISA Kit

Mouse HbA1c(Glycated Hemoglobin A1c) ELISA Kit

Regular price $458.00 USD
Regular price Sale price $458.00 USD
Sale Sold out
Shipping calculated at checkout.
Size
Sensitivity
Detection range

Mouse HbA1c(Glycated Hemoglobin A1c) ELISA Kit

Overview

Product name: Mouse HbA1c(Glycated Hemoglobin A1c) ELISA Kit
Reactivity: Mouse
Alternative Names: Glycosylated Hemoglobin; Hemoglobin A1c; Hb1c; HbAIc; HbAIc
Assay Type: Competitive Inhibition
Sensitivity: 4.55 μg/mL
Standard: 1000 μg/mL
Range: 15.63-1000 μg/mL
Sample Type: plasma and erythrocyte lysates
Assay Length: 2h
Research Area: Diagnosis of diabetes mellitus
Test principle: This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with Mouse HbA1c protein. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Mouse HbA1c. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Mouse HbA1c in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (ng/mL) OD Corrected OD
1000.00 0.182
500.00 0.387
250.00 0.715
125.00 0.933
62.50 1.225
31.25 1.587
15.63 1.813
0.00 2.316

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant HbA1c and the recovery rates were calculated by comparing the measured value to the expected amount of HbA1c in samples.
Matrix Recovery range Average
serum(n=5) 85-97% 91%
EDTA plasma(n=5) 82-94% 88%
Heparin plasma(n=5) 82-94% 91%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of HbA1c and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 87-98% 92-104% 82-95% 86-99%
EDTA plasma(n=5) 92-104% 82-96% 95-102% 87-98%
Heparin plasma(n=5) 88-97% 86-92% 89-97% 91-103%
View full details