1
/
of
1
ELK Biotechnology
SKU:ELK3277
Human PLVAP(Plasmalemma Vesicle Associated Protein) ELISA Kit
Human PLVAP(Plasmalemma Vesicle Associated Protein) ELISA Kit
Regular price
$458.00 USD
Regular price
Sale price
$458.00 USD
Unit price
/
per
Shipping calculated at checkout.
Couldn't load pickup availability
Human PLVAP(Plasmalemma Vesicle Associated Protein) ELISA Kit
Overview
| Product name: | Human PLVAP(Plasmalemma Vesicle Associated Protein) ELISA Kit |
| Reactivity: | Human |
| Alternative Names: | FELS; PV-1; PV1; gp68; Fenestrated-Endothelial Linked Structure Protein; Plasmalemma vesicle protein 1 |
| Assay Type: | Sandwich |
| Sensitivity: | 0.062 ng/mL |
| Standard: | 10 ng/mL |
| Detection Range: | 0.16-10 ng/mL |
| Sample Type: | Tissue homogenates, cell lysates and other biological fluids |
| Assay Length: | 3.5h |
| Research Area: | Metabolic pathway; |
| Test principle: | The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human PLVAP. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human PLVAP. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human PLVAP, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human PLVAP in the samples is then determined by comparing the OD of the samples to the standard curve. |
Standard curve
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.207 | 2.122 |
| 5.00 | 1.657 | 1.572 |
| 2.50 | 1.035 | 0.950 |
| 1.25 | 0.828 | 0.743 |
| 0.63 | 0.496 | 0.411 |
| 0.32 | 0.327 | 0.242 |
| 0.16 | 0.135 | 0.050 |
| 0.00 | 0.085 | 0.000 |

Precision
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Recovery
Matrices listed below were spiked with certain level of recombinant PLVAP and the recovery rates were calculated by comparing the measured value to the expected amount of PLVAP in samples.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-97% | 88% |
| EDTA plasma(n=5) | 80-95% | 87% |
| Heparin plasma(n=5) | 83-95% | 89% |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of PLVAP and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 93-104% | 83-96% | 95-103% | 88-95% |
| EDTA plasma(n=5) | 86-93% | 93-101% | 86-97% | 90-101% |
| Heparin plasma(n=5) | 91-99% | 88-97% | 88-104% | 97-108% |