1
/
of
1
ELK Biotechnology
SKU:ELK2869
Mouse F1+2(Prothrombin Fragment 1+2) ELISA Kit
Mouse F1+2(Prothrombin Fragment 1+2) ELISA Kit
Regular price
$458.00 USD
Regular price
Sale price
$458.00 USD
Unit price
/
per
Shipping calculated at checkout.
Couldn't load pickup availability
Mouse F1+2(Prothrombin Fragment 1+2) ELISA Kit
Overview
| Product name: | Mouse F1+2(Prothrombin Fragment 1+2) ELISA Kit |
| Reactivity: | Mouse |
| Alternative Names: | F1+2 |
| Assay Type: | Sandwich |
| Sensitivity: | 0.114 ng/mL |
| Standard: | 20 ng/mL |
| Range: | 0.32-20 ng/mL |
| Sample Type: | Serum, plasma and other biological fluids |
| Assay Length: | 3.5h |
| Research Area: | Hematology |
| Test principle: | The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Mouse F1+2. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Mouse F1+2. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Mouse F1+2, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Mouse F1+2 in the samples is then determined by comparing the OD of the samples to the standard curve. |
Standard curve
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 20.00 | 1.990 | 1.922 |
| 10.00 | 1.542 | 1.474 |
| 5.00 | 1.043 | 0.975 |
| 2.50 | 0.780 | 0.712 |
| 1.25 | 0.509 | 0.441 |
| 0.63 | 0.305 | 0.237 |
| 0.32 | 0.223 | 0.155 |
| 0.00 | 0.068 | 0.000 |

Precision
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Recovery
Matrices listed below were spiked with certain level of recombinant F1+2 and the recovery rates were calculated by comparing the measured value to the expected amount of F1+2 in samples.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 84-97% | 90% |
| EDTA plasma(n=5) | 95-107% | 101% |
| Heparin plasma(n=5) | 77-95% | 86% |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of F1+2 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-94% | 87-103% | 97-105% | 79-90% |
| EDTA plasma(n=5) | 85-92% | 96-103% | 87-98% | 95-102% |
| Heparin plasma(n=5) | 78-93% | 96-102% | 95-103% | 89-97% |