Overview
| Product name: | Rat TRACP-5b(TartrateResistant Acid Phosphatase 5b) ELISA Kit |
| Reactivity: | Rat |
| Alternative Names: | ACP5;TRAP5b;SPENCDI |
| Assay Type: | Sandwich |
| Sensitivity: | 0.38 ng/mL |
| Standard: | 40 ng/mL |
| Detection Range: | 0.63-40 ng/mL |
| Sample Type: | serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids |
| Assay Length: | 3.5h |
| Research Area: | Immunology;Cardiovascular |
| Test principle: | The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Rat TRACP-5b. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Rat TRACP-5b. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Rat TRACP-5b, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Rat TRACP-5b in the samples is then determined by comparing the OD of the samples to the standard curve. |
Standard curve
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 40.00 | 2.035 | 1.941 |
| 20.00 | 1.593 | 1.499 |
| 10.00 | 1.128 | 1.034 |
| 5.00 | 0.826 | 0.732 |
| 2.50 | 0.614 | 0.520 |
| 1.25 | 0.352 | 0.258 |
| 0.63 | 0.194 | 0.100 |
| 0.00 | 0.094 | 0.000 |

Precision
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Recovery
Matrices listed below were spiked with certain level of recombinant TRACP-5b and the recovery rates were calculated by comparing the measured value to the expected amount of TRACP-5b in samples.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-95% | 87% |
| EDTA plasma(n=5) | 85-97% | 91% |
| Heparin plasma(n=5) | 78-95% | 86% |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of TRACP-5b and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 95-103% | 87-103% | 91-102% | 85-92% |
| EDTA plasma(n=5) | 89-96% | 90-98% | 87-98% | 92-104% |
| Heparin plasma(n=5) | 85-95% | 82-91% | 85-97% | 89-102% |