1
/
of
1
ELK Biotechnology
SKU:ELK9697
Rabbit F1+2(Prothrombin Fragment 1+2) ELISA Kit
Rabbit F1+2(Prothrombin Fragment 1+2) ELISA Kit
Regular price
$458.00 USD
Regular price
Sale price
$458.00 USD
Unit price
/
per
Shipping calculated at checkout.
Couldn't load pickup availability
Rabbit F1+2(Prothrombin Fragment 1+2) ELISA Kit
Overview
| Product name: | Rabbit F1+2(Prothrombin Fragment 1+2) ELISA Kit |
| Reactivity: | Rabbit |
| Alternative Names: | F1+2 |
| Assay Type: | Sandwich |
| Sensitivity: | 1.35 ng/mL |
| Standard: | 200 ng/mL |
| Detection Range: | 3.13-200 ng/mL |
| Sample Type: | serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids |
| Assay Length: | 3.5h |
| Research Area: | Hematology |
| Test principle: | The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Rabbit F1+2. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Rabbit F1+2. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Rabbit F1+2, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Rabbit F1+2 in the samples is then determined by comparing the OD of the samples to the standard curve. |
Standard curve
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 200.00 | 2.135 | 2.054 |
| 100.00 | 1.608 | 1.527 |
| 50.00 | 1.196 | 1.115 |
| 25.00 | 0.875 | 0.794 |
| 12.50 | 0.527 | 0.446 |
| 6.25 | 0.387 | 0.306 |
| 3.13 | 0.156 | 0.075 |
| 0.00 | 0.081 | 0.000 |

Precision
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Recovery
Matrices listed below were spiked with certain level of recombinant F1+2 and the recovery rates were calculated by comparing the measured value to the expected amount of F1+2 in samples.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 81-97% | 89% |
| EDTA plasma(n=5) | 80-95% | 87% |
| Heparin plasma(n=5) | 82-95% | 88% |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of F1+2 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 95-106% | 87-101% | 82-97% | 95-102% |
| EDTA plasma(n=5) | 83-96% | 97-103% | 88-103% | 86-92% |
| Heparin plasma(n=5) | 91-99% | 95-102% | 79-91% | 87-96% |
Share