ELK Biotechnology

SKU:ELK9808

Sheep IL2(Interleukin 2) ELISA Kit

Sheep IL2(Interleukin 2) ELISA Kit

Regular price $376.00 USD
Regular price Sale price $376.00 USD
Sale Sold out
Shipping calculated at checkout.
Sensitivity

Sheep IL2(Interleukin 2) ELISA Kit

Overview

Product name: Sheep IL2(Interleukin 2) ELISA Kit
Reactivity: Sheep
Alternative Names: IL-2;TCGF; Lymphokine; T-Cell Growth Factor; Aldesleukin
Assay Type: Sandwich
Sensitivity: 5.9 pg/mL
Standard: 1000 pg/mL
Detection Range: 15.63-1000 pg/mL
Sample Type: serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Length: 3.5h
Research Area: Cytokine;Infection immunity;
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Sheep IL2. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Sheep IL2. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Sheep IL2, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Sheep IL2 in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (ng/mL) OD Corrected OD
1000.00 2.204 2.111
500.00 1.587 1.494
250.00 1.106 1.013
125.00 0.826 0.733
62.50 0.585 0.492
31.25 0.326 0.233
15.63 0.236 0.143
0.00 0.093 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant IL2 and the recovery rates were calculated by comparing the measured value to the expected amount of IL2 in samples.
Matrix Recovery range Average
serum(n=5) 84-97% 90%
EDTA plasma(n=5) 93-101% 97%
Heparin plasma(n=5) 80-95% 87%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of IL2 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 82-96% 87-98% 89-98% 95-103%
EDTA plasma(n=5) 99-106% 78-92% 85-105% 92-104%
Heparin plasma(n=5) 91-103% 87-98% 92-99% 86-93%